From: Monona Rossol <0000030664c37427-dmarc-request**At_Symbol_Here**LISTS.PRINCETON.EDU>
Subject: Re: [DCHAS-L] Estimating Teaching Lab Occupancy
Date: Sat, 9 Dec 2017 09:11:43 -0500
Reply-To: ACS Division of Chemical Health and Safety <DCHAS-L**At_Symbol_Here**PRINCETON.EDU>
Message-ID: 1603b9e1212-1716-2b09f**At_Symbol_Here**webjas-vab006.srv.aolmail.net
In-Reply-To <00ef01d3707a$6cfbaaf0$46f300d0$**At_Symbol_Here**twc.com>
Excellent point, Bruce
... you can't rely on NFPA for occupational exposure control, their emphasis is fire prevention.
I had to fight years to get the scope of NFPA 1126, Pyrotechnics before a Proximate Audience, to even acknowledge in it's scope that it did NOT cover the health effects of smoke from the pyro.
1126 deals only with safety and fire issues. The only comment about the smoke was that it shouldn't be so thick that it obscures the exit signs! But by that time, you can be seriously in excess of particulate, toxic gases and toxic metal occupational standards. Many pyro effects are based on lead-styphnate and other toxic metal compounds.
Yet people in my biz often follow 1126 and declare the effect safe in all respects. And people who tell me their theater or art studio is safe because the fire marshal went through and said everything was fine need to get a grip!
OK, class, now that brings us back to the safety standards for classroom demonstrations:
WE REALLY NEED to look at the exposure issues in some of these demos. If we are burning metal compounds to see the pretty colors or volcano effects, maybe we should monitor exposures to the metals and find out where they end up when they settle. Copper, manganese, chromium, and barium have TLVs. Some metals are known sensitizers and asthmagens.
And a volcano made with a dichromate clearly exposes spectators to chromium and creates an indoor superfund area.
Monona Rossol, M.S., M.F.A., Industrial Hygienist
President: Arts, Crafts & Theater Safety, Inc.
Safety Officer: Local USA829, IATSE
181 Thompson St., #23
New York, NY 10012 212-777-0062
-----Original Message-----
From: Bruce Van Scoy <bvanscoy**At_Symbol_Here**TWC.COM>
To: DCHAS-L <DCHAS-L**At_Symbol_Here**PRINCETON.EDU>
Sent: Sat, Dec 9, 2017 6:11 am
Subject: Re: [DCHAS-L] Estimating Teaching Lab Occupancy
Laurence,
Prudent Practices was significantly revised in 2011 and is still available free from NAP.edu and maintained a basic requirement (see section 9.B or pg. 213 or 9.C/pg. 219) for ventilation. But Prudent Practices is to basic to be useful for design and you can't rely on NFPA for occupational exposure control, their emphasis is fire prevention.
I strongly recommend following ACGIH's Industrial Ventilation: A Manual of Recommended Practice for Design, 29th Edition ALONG with Industrial Ventilation: A Manual of Recommended Practice for Operation and Maintenance, also 29th edition.
I would also strongly recommend that if an architect acts as specified below, DOCUMENT, with citations AND copies to your attorney at every step-.
And, oh by the way, if one happens to mention "value engineering" ensure that EHS has to sign off on ALL design changes in the contract language UP FRONT. Otherwise, a design may turn into a bait and switch with significant differences between as designed v. as built. Also, make sure they are keeping up with the as-built drawings. You could find yourself looking at specifications/prints later that are totally inaccurate.
Just my personal opinions,
BruceV
This citation is probably dated but it is a good starting point. the book has a chapter on laboratory design.
Hi DCHAS colleagues,
Does anyone know the best way to estimate occupancies for college teaching labs based on square footage and the exact sources for those estimates? I've seen several sources state 50 ft2 and cite the NFPA but I'm having trouble figuring out how to locate that specific information in the fire code.
Also, does anyone know if that refers to gross or net (i.e. after benches/hoods/etc.) square footage?
If you want some context- I've include all the drama-laden details below.
Instructor, Pasadena City College
Details for context (Warning: drama-ahead!):
We're getting a new science building to replace our old building that was condemned for seismic issues. We just found out a month ago that the plans for the building were somehow finalized at the state chancellor's office without anyone's knowledge and that the submitted plans were designed in-house by someone in facilities rather than an actual architect. Based on our analysis, the new building will drop our lab enrollment capacity to almost 75% of what we can currently manage, which is 70% of what we where historically offering before our old building was condemned. (There are other issues, including that its five stories with one elevator.) We've been told very firmly that we can not make any changes to the building without losing our spot in the funding queue. (We're fine with this, the district board is not.)
When we talked to the architects about the square footage issue, they claimed that egress was the only issue for teaching lab occupancy and that they were unfamiliar with the 50 ft2/student value we were citing. [The architects were also rather mysteriously chosen since they didn't appear as one of the finalists selected by the committee. We've found several news articles about lawsuits involving code violations by this firm.)
---
For more information about the DCHAS-L e-mail list, contact the Divisional secretary at
secretary**At_Symbol_Here**dchas.org
Follow us on Twitter **At_Symbol_Here**acsdchas
---
For more information about the DCHAS-L e-mail list, contact the Divisional secretary at secretary**At_Symbol_Here**dchas.org
Follow us on Twitter **At_Symbol_Here**acsdchas
Previous post | Top of Page | Next post